Firmware Upgrade Tool for G460 Gas Detectors

 
To upgrade the firmware in a G460 you must have a charging cradle and a GfG data cable. The data cable is distributed with the
G460 data download software kit.
Please refer to G460 data download software page for details.

Below are the instructions for downloading the G460 firmware Upgrade Tool.
Note: Depending on the computer configuration an administrative log-in may be required.
 
 
G460 Firmware
 

File:

G460v342_Update.exe

Document Type:

Executable (EXE)

Version:

3.42

Release Date;

26 May 2011

File Size:

857 KB

 
 
Requirements
 

Computer:

Any Pentium or equivalent with Windows 2000 or later with one available USB port.

GfG Hardware:

Charging cradle and G460 data cable (distributed with GfG data download kit for G460)

 
 
  1. Download the G460 firmware Upgrade Tool by clicking on the file name and save it.
    Make a note of the location to where it is saved.

  2. Make sure the G460 is placed in a charging cradle and that the GfG USB cable is connected between the cradle and the computer.
    If the "New Hardware Found" wizard starts when the cable is connected to the computer then you must first complete the driver installation using only GfG Drivers. DO NOT allow windows to install drivers automatically.
    See the G460 data download software page for details.

  3. Using Windows Explorer (My Computer) navigate to the folder where the downloaded file was stored.

  4. Double click on the file to start the Upgrade Tool.
    The program does not require installation and can be executed directly.

  5. Select the GfG COM port (this is usually the last one on the list) and click connect
    If not you selected the wrong COM port or otherwise do not have a connection or the drivers were installed incorrectly the connection will fail.

  6. Click the [Start Update] button.
    The G460 should chirp and a progress bar on the screen should indicate to state of the process.
    If not you selected the wrong COM port or otherwise do not have a connection or the drivers were installed incorrectly.
    Do not disconnect the cable or remove the G460 until the process is complete (approximately 60 seconds).

  7. Once the process is complete the instrument will chirp again and restart.
    The new firmware version will be displayed on the first screen as the G460 is starting up.

 
 
 